The miniTOF® II mass spectrometer system provides fast answers to mass analysis questions at low cost. Complete with TOF, heated ion source, vacuum chamber and pumps, all necessary detection and control electronics, Windows-based software and power supplies for analysis of positive ion spectra, the miniTOF® II system is suitable for both laboratory and for OEM uses. NIST library search capability is available.
Applications include fast GC-MS and gas monitoring and analysis. Specifications include mass resolution of better than 1 Da for the mass range from 1 to over 1200 Da, cycles rates to 20 kHz, and ppm level sensitivity. System dimensions are approximately 22”(D) x 12”(W) x 20” (H). Complete systems are available from $50K. Quantity discounts are available.
MECHANICAL
Comstock
miniTOF® II is comprised
of a vacuum vessel with a linear TOF spectrometer inside, a PC system meeting
today’s standards, an external power supply box, and vacuum pumps. Vacuum seals
are Viton O-rings. (Seals of other
materials may be used. Contact
factory.)
Location of Major
Components
The
spectrometer itself is attached to the inside of the front panel of the vacuum
vessel for convenient removal, and a small satellite power supply is on the
other (air) side past electrical feedthroughs. The main power supply is in a metal box,
normally placed underneath the vacuum vessel. The Time-to-Digital Converter, control
firmware, and digital as well as analog interfaces are on a single specially
designed PC card and housed in the PC tower. A turbomolecular drag vacuum pump(70
l/s) is also attached to the vacuum
vessel. An oil free diaphragm foreline pump is normally placed on the
floor. A port is on the vessel for
an optional vacuum gauge.
Spectrometer dimensions are
approximately 22” (D) x 12” (W) x 20” (H), excluding PC, CRT and
pump.
ELECTRONICS
Power Supplies Generates all the voltages
and currents necessary to operate the spectrometer. A number of these voltages are singly or
doubly regulated for the precision operation of the
system.
Time-to-Digital
Converter Resides on a PCI card housed
in the PC. It converts the time
elapsed between repeller time and detection at the end of the flight into a
digital form, when the information is transferred from the clock to a FIFO
memory. Multiple stops are allowed
to each start. One count is
recognized for each of the stops regardless of the amplitude of the signal from
the detector so long as it exceeds the discriminator level. The time data stored in the FIFO memory
is transferred to the PC memory on firmware command. The time resolution of the clock (bin width)
is 10 nanosecond. Time count is 32 bit long. The event count capacity used in the PC
is 64 bits.
PC
Interface/Controls Digital and analog I/O
interface ports are available on the PC card, most of which are used for the
systems purposes. The use of extra
ports for additional functions are possible but requires special engineering and
programming. Contact
factory.
Safety
Interlocks In order to safeguard the
spectrometer components inside the vacuum vessel, a special sensor in the
turbomolecular pump signals the pressure inside the vessel. Voltages are disabled until a proper
vacuum is reached.
SPECTROMETER
General The spectrometer is a
linear, orthogonal injection, electron impact ionization Time-of-Flight
spectrometer. The entire instrument, from sample intake to ion detector and gas
exhaust is carefully designed and carefully machined for optimum space and time
focusing. Standard sample inlet is
an undeactivated 14 mil OD capillary in a Cajun connector mounted on the front
panel of the vacuum vessel that provides a pressure of approx 5x10^-5 Torr to
the TOF. The capillary is a standard GC
capillary. (Other sample inlet arrangements are possible. Contact factory.) The flight path is about 40 cm from
repeller to detector. The detector
is a discreet dynode type for sensitivity and ruggedness.
Operation Operating parameters are set on the PC screen and communicated to the power supplies via the interface board. Filament current and detector bias are monitored back on the PC screen so that the operator can be assured that there is no significant deviations.
Upkeep The entire spectrometer is
mounted on the vacuum side of the front panel to ensure easy access. Electron emitting filament is positioned
by two screws that are easily accessible.
It takes less than 30 minutes from the time the vacuum pumps are turned
off until they are turned back on with a new filament
inside.
PERFORMANCE*
Mass Resolution Typical M/dM resolution as it is tested at the factory is about 1,200.
Mass Range 1 ~ 1200 nominal.
Stability Peak shift < 0.02% over 24 hours.
Peak broadening not observed over 24 hours.
Minimum Detection Limit 5 ppm without pre-concentration. The limit is a function of measurement time (= number of sweeps) and the gas mixture injected.
Dynamic Range The dynamic range is a function of measurement time and discriminator setting, among other parameters. Due to the nature of TDC counting, the dynamic range, defined as the maximum count that can be stored in the system divided by the least significant count, can be an extremely large number. The maximum event counts (the vertical scale on the spectrum display) is 2^64.
Mass
Calibration Carried out on quadratic
model based on three or more data points.
Single set of coefficients over the mass range. Calibration coefficients could change
significantly when detector voltage is changed. Operating software issues a warning in a
pop up window when it sees an old mass calibration. Operator defines how old is old. When
the software does not see calibration performed for that voltage, it refuses to
convert the time scale (horizontal) to mass scale.
SOFTWARE
Labview®
The programs provided with the system are written
in LabView® language under Windows OS.
They cover set up of operational parameters, data acquisition, data
handling between TDC and PC and inside PC, display, and mass calibration. A hook
to the search engine for NIST’s MS library as well as access to the library data
itself are available. If an additional function is needed, contact
factory.