High Resolution - High Transmission Reflectron Time-of-Flight Mass Spectrometers


![]()
Comstock's 200 series reflectron Time-of-Flight mass spectrometers are the state of the art in commercial Time-of-Flight instrumentation.
The combination of Comstock's proprietary second order spatial focussing sources for EII, LI, MALDI, and LDI applications and the second order energy focussing ion mirror results in superb mass resolution. Under the conditions of electron impact ionization of room temperature gases, the mass resolution is 5000 m/dm (FWHM) for molecular gases and 9000 m/dm (FWHM) for atomic gases.
The gridless design and Comstock's proprietary dispersion correcting lensing of the ion mirror results in nearly perfect transmission of ions through the ion mirror. The transmission, determined using ion counting techniques, is 92.7 +/- 0.5 %.
Please inspect our test data from our RTOF-210 equipped with EII.
![]()
RTOF-210 Electron Impact Ionization Spectra
CO and N2, Demonstrates the resolving power of Comstock's RTOF-210 reflectron spectrometer by showing the separation of N2 (m/z=28.00614 amu) and CO (m/z=27.99491 amu).
Perfluorophenanthrene, Calibration compound, includes peak table showing mass accuracy.
Xenon, Shows resolution of EII-RTOF-210 averaged spectrum with atomic gas, abundance distribution of xenon isotopes, includes simulated spectra of xenon with isotopes.
![]()